Fast & Easy
Fast sterilization cycles
and easy to use
Fast sterilization cycles
and easy to use
Minimizes damage to delicate
and high-value items
Multi-use sterilant cartridge
for cost efficiency
Non-oil system
for easy maintenance
| Category | Details | Unit |
|---|---|---|
| Dimensions | 542 (W) × 418 (D) × 308 (H) | mm |
| Chamber Capacity | 470 (W) × 185 (D) × 90 (H), 8 | mm, l |
| Chamber Diagonal Length | 510 | mm |
| Product Weight | 30 | kg |
| Sterilization Process | Sterilization Time (Standard Mode): 18 ±2 | min |
| Sterilization Time (Advanced Mode): 31 ±2 | min | |
| Sterility Assurance | SAL 10−6 validated for medical device sterilization safety | - |
| Sterility Assurance Level (99.9999%) | - | |
| Safety Devices | Physical: Door sensor and vacuum-based door locking prevention | - |
| Electrical: Fuse installed for overcurrent protection | - | |
| Vacuum | Oil-free vacuum pump | - |
| Minimum Pressure ≤ 30 | Torr | |
| Rate of Pressure Change ≤ 6 | Torr/min | |
| Electrical Characteristics | Rated Voltage: 220 | VAC |
| Rated Frequency: 50 / 60 | Hz | |
| Power Consumption: 1.8 | kW | |
| Fuse: 250V, 10A | - | |
| Process Variables | Sterilization Temperature: 30 ~ 60 | ℃ |
| Pressure before Sterilant Injection ≤ 30 | Torr | |
| Sterilant Diffusion Pressure: 25 ~ 80 | Torr | |
| Sterilant | Hydrogen Peroxide (H₂O₂ 58.0% ~ 59.5%) | - |
| Single-channel Lumen Test | Ø0.7 ×500 mm Stainless steel, 3EA | - |
STERLOAD lite+
| Mode | Cycle | Cycle time |
|---|---|---|
| Standard | 30 | 18 min |
| Advanced | 30 | 31 min |